NEWMEEN
Atomic Absorption Spectrophotometer | Flame & Graphite Furnace AAS TAS990
Engineered for elemental trace quantification across environmental monitoring, food testing, geochemistry, pharmaceuticals, and metallurgical sectors
High-Throughput Optical Bench and Automated Lamp Configuration
The optical assembly features a high-throughput Czerny-Turner monochromator equipped with an 1800 lines/mm holographic diffraction grating, delivering exceptional wavelength accuracy of ±0.25 nm and resolution of 0.2 nm ± 0.02 nm
Dual Atomization with Transversely Heated Graphite Furnace
The electrothermal atomization system utilizes advanced transverse graphite furnace heating, ensuring uniform isothermal heating along the optical path
Integrated Safety Interlocks and Flexible Analytical Accessories
Operational safety is ensured by real-time hardware interlocks that continuously monitor oxidant support pressure, fuel flow rates, drain liquid levels, flame sensors, and burner head positioning switches
Technical Specifications
| Parameter Category | Specific Technical Parameter | Specification / Performance Standard |
| Optical System | Wavelength Coverage | 190 nm – 900 nm |
| Monochromator Geometry | Czerny-Turner Configuration | |
| Grating Density | 1800 lines/mm | |
| Spectral Bandwidth | 0.1 nm, 0.2 nm, 0.5 nm, 1.0 nm, 2.0 nm (automated 5-slit selection) | |
| Wavelength Accuracy | ±0.25 nm | |
| Wavelength Reproducibility | ≤ 0.15 nm | |
| Spectral Resolution | 0.2 nm ± 0.02 nm | |
| Baseline Stability | ±0.004 Abs / 30 min | |
| Background Correction | Deuterium Arc (1.0 Abs) & Self-Reversal (1.0 Abs, > 55 times) | |
| Detector Type | Photomultiplier Tube (PMT) | |
| Lamp Turret Mechanism | Motorized 8-position hollow cathode lamp turret with auto-alignment | |
| Flame Analysis | Applicable Flame Chemistries | Air/Acetylene; Nitrous Oxide/Acetylene; Air/Propane (LPG) |
| Characteristic Concentration | Cu < 0.04 μg/mL / 1% (Air/Acetylene); Ba < 0.15 μg/mL / 1% (N₂O/Acetylene) | |
| Detection Limit | Cu ≤ 0.006 μg/mL | |
| Precision (Repeatability) | Cu < 1.0% RSD; Ba < 1.0% RSD | |
| Analytical Sensitivity | Cu 2 μg/mL Absorbance > 0.28 Abs | |
| Burner Head Material | Corrosion-resistant Metal Titanium Alloy | |
| Nebulizer Assembly | High-efficiency glass nebulizer (acid-proof option available) | |
| Burner Adjustment | Automatic height adjustment and automatic changeover between flame and furnace | |
| Graphite Furnace | Atomization Temperature Range | Ambient to 2650°C |
| Detection Limit | Cd ≤ 0.01 ng/mL | |
| Precision (Repeatability) | Cu < 3.0% RSD; Cd < 4.0% RSD | |
| Heating Architecture | Transverse horizontal graphite furnace heating | |
| Temperature Control Method | Real-time voltage feedback control for drying and ashing stages | |
| Temperature Precision | < 1% error | |
| Operational & Safety | Safety Interlock Systems | Support gas pressure, fuel line shutoff, drain trap liquid level, flame sensor, emergency kill button |
| Furnace Protection | Argon gas flow monitoring, cooling water flow cut-off, tube fracture sensor | |
| Software Platform | Dedicated Windows software featuring multi-order calibration and automated curve fitting |
Frequently Asked Questions (FAQ)
1. How does the system transition between flame and graphite furnace modes?
The instrument features automated mechanical positioning that shifts between flame and graphite furnace atomizers via software commands without manual optical realignment
2. What are the operational advantages of transverse graphite furnace heating?
Transverse heating creates an isothermal temperature profile throughout the graphite tube, eliminating cold ends, reducing chemical interferences, and preventing sample memory effects
3. What background correction options are available for complex sample matrices?
The analyzer provides both a Deuterium Arc lamp and high-speed Self-Reversal background correction, delivering over 55-fold correction capability for high-absorbance interferences
4. What elements can be determined using the dedicated hydride generation accessory?
The hydride generator is optimized for ultra-trace quantification of volatile hydride-forming elements including arsenic, selenium, lead, bismuth, antimony, tellurium, tin, germanium, and cold vapor mercury
5. What automated safety interlocks protect personnel during flame atomization?
Integrated safety systems monitor oxidant support pressure, drain liquid trap sensors, burner installation interlocks, flame status monitors, and an emergency flame kill switch
6. What protective measures prevent damage to the graphite furnace module?
The furnace controller monitors argon carrier gas pressure and external cooling water flow rates, automatically cutting power if pressure drops or if a graphite tube fractures
7. How does the flame autosampler handle low-concentration samples?
The flame autosampler supports automated routine calibration and sample introduction alongside automated online pre-concentration and elution for low-level target analytes.
8. What capabilities are integrated into the graphite furnace autosampler?
The autosampler executes automated multi-point standard curve generation, matrix modifier additions, intelligent sampling depth adjustments, and air-barrier liquid isolation to prevent cross-contamination.
9. Why is a dedicated temperature-controlled water chiller required for graphite furnace analysis?
A circulating water chiller ensures high-efficiency cooling, maintains uniform furnace temperature repeatability, and prevents internal mineral scaling caused by untreated tap water.
10. What are the benefits of using an oil-free air compressor with the flame unit?
An oil-free compressor delivers clean, oil-free, dry compressed air to the burner, preventing carbon contamination in the mixing chamber and avoiding erratic flame baseline fluctuations.
11. What is the capacity and operating function of the automated lamp turret?
The motorized turret accommodates 8 hollow cathode lamps simultaneously, allowing automated optical line peaking, position optimization, and automated element warm-up sequences
12. Which flame chemistry types are supported for refractory element analysis?
The titanium burner supports standard Air/Acetylene, high-temperature Nitrous Oxide/Acetylene for refractory elements like barium and aluminum, and Air/Propane (LPG) mixtures