Laboratory equipment BioProcess bags

NEWMEEN

M6 Microwave Digestion System

reviews
items available
Add inquiry cart Inquire Now

M6 microwave digestion system delivers the sample preparation solution according to your need. With high-quality material and innovative solution as rotor and vessel position recognition, M6 helps to achieve reliable trace element analysis result due to precise temperature control ability Status light visualizes the program running status and the LCD screen displays the information as each vessel's temperature via histogram. The flexible and extendable rotor allows different application such as acid digestion, solvent extraction and other sample preparation process of various sample types such as soil, food, drugs, materials etc.,.

M6 Microwave Digestion System


  • M6 Microwave Digestion System
    Compatible for 3 different rotors
  • M6 Microwave Digestion System
    Auto recognition and full vessel monitor
  • M6 Microwave Digestion System
    160 sample per day



Guaranteed user friendliness

  • Pre-loaded digestion methods

  • Intuitive software interface

  • Convenient switch between rotors

  • Multiple safety features in the hardware

    Guaranteed digestion performance

    • Complete solutions to samples of diverse background

    • Precise dual temperature/pressure control

    • Max batch time (including cooling) NMT 60 min

      Guaranteed Return on Investment

      • Maximized sample throughput (160 per shift)

      • Reduced average acid consumption

      • Minimized consumables and extended vessel lifetime


Oven Body

5-year corrosion-free warranty

  • 316L Stainless Steel oven body construct

  • Seamless Laser Soldering

  • 5-Layer Premium Teflon Coating


    Microwave Emitter

    Proprietary microwave focusing

    • Optimized dual-magnetron positioning

    • Creative microwave guiding technology to streamline the energy density and distribution

    • Automatic microwave power adjustment to achieve target temperature

    • Delivers rapid, safe, and reproducible batch digestions for as many as 40 samples


    Enhanced Safety Design

    Buffered safety door concept

    • Floating Safety Lock

    • Automatic depressurize and re-close

    • Secure microwave emission upon fully closed door



      Vertical Rotor Release and Engagement

      Rotor can be vertically raised or lowered by the motion of the mechanically interlocked safety door

      Rotor handling is greatly facilitated

      Operational Mechanism:

      Rotor raised when safety door is opened to a ≥ 90°

      Rotor lowered when safety door is opened to a< 90°


      Microwave Digestion Rotor Series

      • Ultra High Performance Rotor UHP10:

        • Unrivalled temperature and pressure tolerance in microwave digestion.

        • Tailored digestion solution ideal for the toughest sample types such as graphite, activated carbon, zircon, alumina, etc.

        • Guaranteed performance durability and reliability.

      • High Performance Rotor HP16:

        • Greater temperature and pressure tolerance for tough sample matrices.

        • Max 16 samples in a single batch.

        • Powered by Rtemp Mid-IR technology.

        • Contactless and real-time T/P control in full vessel set.

      • High Throughput Rotor GT-400:

        • Max 40 samples in a single batch.

        • Powered by Rtemp Mid-IR technology.

        • Contactless and real-time T/P control in full vessel set.


      RotorUHP10HP16GT-400
      No. of vessels101640
      Vessel materialTFMTFMTFM
      Protective sleeve materialReinforced PEEKReinforced PEEKReinforced PEEK
      Vessel volume100mL100mL60mL
      Max withstanding temperature330°C330°C330°C
      Max withstanding pressure150bar150bar120bar


      FeatureSpecification Details
      Touch screen7" High def colour LCD
      Microwave ovenIndustrial grade resonance oven
      Oven body316L Stainless Steel
      AC input220-240V/50Hz, 16A
      Total power3200W
      Microwave frequency2450MHz
      Microwave emission modeUnpulsed, ±1W
      Microwave control modePID control
      Ambient temperature5-40°C
      Ambient humidity15%-80%RH
      Dimension of mainframe665(H) × 546(W) × 590 (D) mm
      Net weight70Kg


      FAQs

      1. What materials are used in the construction of the oven body?

      The oven body is constructed from 316L Stainless Steel featuring seamless laser soldering and a 5-layer premium Teflon coating for maximum corrosion resistance.

      2. What is the maximum sample capacity for a single batch?

      Depending on the rotor selected, the system can process up to 40 samples simultaneously using the GT-400 high-throughput rotor.

      3. How does the system control the microwave emission?

      The system utilizes PID control and an unpulsed emission mode with a precision of ±1W to ensure stable heating.

      4. What are the safety features of the door design?

      The system features a buffered safety door concept with a floating safety lock and automatic depressurization and re-close functions.

      5. What is the maximum pressure the digestion vessels can withstand?

      The Ultra High Performance (UHP10) and High Performance (HP16) rotors are designed to withstand pressures up to 150 bar.

      6. Does the system provide real-time temperature and pressure monitoring?

      Yes, it utilizes Rtemp Mid-IR technology for contactless, real-time temperature and pressure control across the full vessel set.

      7. What is the benefit of the vertical rotor release mechanism?

      The rotor can be vertically raised or lowered via the mechanically interlocked safety door, which greatly facilitates sample handling.

      8. Can this system handle "tough" sample matrices like graphite or alumina?

      Yes, the UHP10 rotor is specifically tailored for the toughest sample types, including graphite, activated carbon, zircon, and alumina.

      9. What are the power and electrical requirements for installation?

      The system requires an AC input of 220-240V/50Hz (16A) with a total power consumption of 3200W.

      10. What material is used for the digestion vessels and sleeves?

      The vessels are made of TFM, while the protective sleeves are constructed from Reinforced PEEK for high-strength performance.

      11. What is the total processing time per batch?

      The system provides high-performance digestion with a maximum batch time, including cooling, of no more than 60 minutes.

      12. How does the dual-magnetron positioning improve performance?

      The optimized dual-magnetron positioning and proprietary microwave guiding technology streamline energy density and distribution for reproducible results.


Recent Reviews
+